Chronoamperometry/Chronocoulometry
Introduction
Chronoamperometry (CA) and chronocoulometry (CC) have the same potential wave form - the potential step - which is one of the simplest potential wave forms. As shown below (F1), the potential is changed instantaneously from the Initial Potential to the First Step Potential, and it is held at this value for the First Step Time. This is a single potential step experiment. In a double potential step experiment, the potential is changed to the Second Step Potential after the First Step Time, and it is then held at this value for the Second Step Time. In CA, the current is monitored as a function of time, whereas in CC, the charge is monitored as a function of time. It is important to note that the basic potential step experiment on the epsilon is CA; that is, during the experiment, the current is recorded as a function of time. However, after the experiment, the data can also be displayed as charge as a function of time (the charge is calculated by integrating the current). Hence, chronocoulometry data can be obtained. CA is a standard technique on the epsilon.CA is different from other constant potential techniques (constant potential electrolysis (CPE) and DC potential amperometry (DCA)) in that the time scale of CA is shorter (milliseconds and seconds) than those of BE and DCA (seconds and minutes).
Figure 1. Potential wave form for chronoamperometry and chronocoulometry.
Setting Up a Chronoamperometry/Chronocoulometry Experiment
As shown above, five parameters are used in the epsilon software to define the potential wave form for CA.- Initial Potential
- First Step Potential (E)
- Second Step Potential (E)
- First Step Time
- Second Step Time
Figure 2. Change Parameters dialog box for chronoamperometry/chronocoulometry.
- Potential values are entered in mV, and time values in ms or s (select using Time Units).
- If the Apply Open Circuit Potential for Initial E box is checked, then the open circuit potential will automatically be measured and used as the Initial Potential.
- If the Run - External Trigger box is checked, the experiment is started from an external device using the Start In back-panel connection.
- When the experiment is started, the cell is held at the Initial Potential for the number of seconds defined by the Quiet Time.
- The experiment can be run on a hanging mercury drop electrode (i.e., a single drop is used for the entire experiment) using a BASi CGME by selecting CGME SMDE Mode from Cell Stand / Accessories in the Setup / Manual Settings (I/O) dialog box.
- A rotating disk experiment can be run using a BASi RDE-2 by selecting RDE-2 from Cell Stand / Accessories in the Setup / Manual Settings (I/O) dialog box and entering the required Rotation Rate under RDE2 Rotation in the Cell dialog box.
- There are two gain stages for the current-to-voltage converter. The default values of these stages that are used for a given current Full Scale value are determined by the software. However, they can be adjusted manually using the Filter / F.S. dialog box. This dialog box is also used to change the analog Noise Filter Value settings from the default values set by the software.
- The default condition of the cell is that the cell is On (i.e., the electronics are connected to the electrodes) during the experiment, and is Off between experiments. However, the potential can be switched On between experiments using the Cell dialog box. HOWEVER, THIS OPTION SHOULD BE USED WITH CAUTION SINCE CONNECTING OR DISCONNECTING THE ELECTRODES WHEN THE CELL IS ON CAN RESULT IN DAMAGE TO THE POTENTIOSTAT, THE CELL, AND/OR THE USER!
- A series of identical experiments on the same cell can be programmed using the MR (Multi-Run) option.
- Clicking Exit will exit the dialog box without saving any changes made to the parameter values. Any changes can be saved by clicking Apply before exiting.
- Range of allowed parameter values:
- Potential = -3275 mV - +3275 mV
- Quiet Time = 0 - 100 s
- Step Time = 1 - 65 s OR 1 - 16000 ms
- Maximum # of points in a step = 1000, 2000, 4000, 8000, 16000
- The Sample Interval is determined by the Step Time and the Maximum # of points in a step, and can only be adjusted by the user indirectly through these latter two parameters. The relationship between these parameters is shown by the equationSample Interval = Step Time/Maximum # of points
However, it should be noted that only certain values are allowed for each of these parameters, as is shown in the table below:
Max. # of points 1000 2000 4000 8000 16000 Sample Interval
Maximum Step Time (/ms) 20 ms 20 40 80 160 320 50 ms 40 81 162 325 650 100 ms 100 200 400 800 1600 200 ms 200 400 800 1600 3200 500 ms 406 812 1625 3250 6500 1 ms 1000 2000 4000 8000 16000 2 ms 2000 4000 8000 16000 5 ms 4062 8125 10 ms 10000
- Once the parameters have been set, the experiment can be started by clicking Run (either in this dialog box, in the Experiment menu, in the pop-up menu, on the Tool Bar, or using the F5 key).
Analysis of the Current Response
Let us consider the effect of a single potential step on the reaction R = O + e-. At potentials well negative of the redox potential (Enr), there is no net conversion of R to O, whereas at potentials well positive of the redox potential (Ed), the rate of the reaction is diffusion-controlled (i.e., molecules of R are electrolyzed as soon as they arrive at the electrode surface). In most potential step experiments, Enr is the Initial Potential, and Ed is the First Step Potential. The advantage of using these two potentials is that any effects of slow heterogeneous electron transfer kinetics are eliminated. In double potential step experiments, (Enr) is often used as the Second Step Potential.Since the current is directly proportional to the rate of electrolysis, the current response to a potential step is a current 'spike' (due to initial electrolysis of species at the electrode surface) followed by time-dependent decay (F4) (due to diffusion of molecules to the electrode surface).
Figure 4. Current-time response for a double-potential step chronoamperometry experiment.
For a diffusion-controlled current, the current-time (i-t) curve is described by the Cottrell equation:
i = nFACD½p-½t -½
where: | n = number of electrons transferred/molecule F = Faraday's constant (96,500 C mol-1) A = electrode area (cm2) D = diffusion coefficient (cm2 s-1) C = concentration (mol cm-3) |
Q = 2nFACD½p-½t ½
Figure 5. Charge-time response for a double-potential step chronocoulometry/chronoamperometry experiment.
Charge is the integral of current, so the response for CC increases with time, whereas that for CA decreases. Since the latter parts of the signal response must be used for data analysis (the finite rise time of the potentiostat invalidates the early time points), the larger signal response at the latter parts for CC makes this the more favorable potential step technique for many applications (in addition, integration decreases the noise level).
The analysis of CA and CC data is discussed elsewhere.
http://www.basinc.com/mans/EC_epsilon/Techniques/ChronoI/ca.html